| Operating Procedures |
|
AC Light On whenever pump is plugged in.
Rate Control Knob Increases or decreases flow rate desired.
ON Powers the pump.
OFF Shuts pump off.
|
|
|
START/HOLD Starts pump rotor to deliver feeding formula and alternately puts unit on HOLD.
HOLD is used to stop fluid flow to:
1. Correct alarm condition (alarm is silenced)
2. Change the flow rate
3. Change pump set
NOTE: IF THE PUMP HAS BEEN PLACED
ON HOLD AND NOT RESTARTED WITHIN 2.5 MINUTES, AN ALARM WILL
SOUND.
NOTE: BATTERY WILL CHARGE
WITH PUMP PLUGGED INTO WALL OUTLET, EVEN WHEN OFF BUTTON HAS BEEN
PRESSED.
|
|
Normal Operations Insert hospital grade plug into AC wall
outlet.
|
|
Battery Operation Unplugging the pump from the AC line outlet will
automatically put the pump on battery power. A new battery, when fully
charged, will operate the instrument for more than 24 hours at 125
milliliters per hour. Approximately 15 minutes prior to battery discharge,
a low battery alarm will occur, (see alarm section). When complete
discharge occurs, the pump will automatically turn itself off.
NOTE: AS THE BATTERIES GET OLDER,
THE TIME FROM LOW BATTERY ALARM TO COMPLETE DISCHARGE MAY BE LESS THAN 15
MINUTES.
The battery will charge
continuously during AC line operation.
To recharge battery, insert
hospital grade plug into AC wall outlet. Battery will automatically begin
charging. The recharge time for a fully discharged battery is
approximately 15 hours.
The battery will charge whenever
the pump is plugged into a wall outlet, even when the OFF button has been
pressed. To be certain that power to all circuits has been disconnected,
the power cord must be unplugged from the wall outlet.
|
|
Beginning Administration Below are general instructions for inserting all
Sherwood Medical Pump Sets into the KangarooŽ 224 Feeding Pump.
1. Press ON.
2. Wait while pump performs
system check.
3. When
flow rate is displayed, set desired rate
in ml/hr by turning rate control knob on side of pump.
|
|
|
|
| 4. Slide
drip chamber into upper drip chamber guide.
|
5. Seat
and secure chamber in lower drip chamber guide. (A).
|
|
6. Remove distal connector
protective cover.
7. Slowly open clamp and fill
entire line with fluid, leaving a few drops of fluid in the drip
chamber.
NOTE:AVOID FILLING THE DRIP
CHAMBER MORE THAN HALF FULL.
8. Close clamp
9. Attach distal connector to
feeding tube.
|
|
|
|
10. Grasp silicone tubing and
gently stretch around rollers. Insert retainer into retainer lock (B).
Thread tubing over tubing guide (C). Secure latch over tubing (D).
- VERIFY THAT THE PUMP SET IS PROPERLY LOADED IN THE PUMP BEFORE
PROCEEDING. AN IMPROPERLY LOADED PUMP SET COULD RESULT IN UNCONTROLLED
FLOW AND POSSIBLE PATIENT INJURY.
11. Open control clamp.
12. Press START/HOLD.
13. Proceed with feeding.
|
| Example of
proper placement of Pump Set in Kangaroo 224 Pump. |
|
|
Back to Table of Contents |
Alarms |
|
The Sherwood Medical KangarooŽ 224
Feeding Pump has been designed to provide audible and visible alarms under
several conditions where proper performance cannot be maintained. When an
alarm occurs, press the START/HOLD button on the pump and perform the
necessary procedures to correct the alarm condition.
NOTE: THE VISIBLE ALARM IS IN THE
FORM OF A MESSAGE SHOWN ON THE DISPLAY. |
Visible Alarm
Definitions |
No Set (no SEt) The feeding set is not installed or is installed improperly.
See Operating Procedures for installation instructions.
Flow Error (FLO Err)
-
The feeding
container is empty. To correct this situation, if more feeding is
required, refill the container with the desired amount of formula and
follow the steps in the Operating Procedures of this
manual.
-
The feeding
container, the feeding set tubing, or the feeding tube itself has become
occluded beyond the pump's capability to pump. Alarm will activate at
approximately 12 psi (8~.7 kPa). To correct this condition, determine
where in the system the occlusion has occurred. Correct occlusion
condition. Press START/HOLD switch. Continue with feeding.
-
The drop
sensors are blocked and will not be able to detect drops as they come
through the drip chamber. To correct this condition, check drop sensors
on inside of upper drip chamber guide and clean off any material that
may be blocking sensors.
-
This alarm is
activated when the battery begins to run down and flow rate accuracy can
no longer be maintained. The low battery alarm occurs approximately 15
minutes before complete discharge. When complete discharge occurs, the
pump automatically turns itself off.
-
To correct this
condition, plug the pump into an AC wall outlet. When the pump is
connected to a wall outlet, it will continue to operate and charge the
battery at the same time. Once the low battery alarm has sounded, it
will take approximately 15 hours to fully recharge.
Hold Error (Hid Err) If the pump is left in the hold mode for approximately
2 1/2 minutes "HOLD ERROR" will alarm.
System Error (Sys Err)
- Rotor is not turning while
pump is in start mode.
NOTE: DOTS WILL FLASH ACROSS TOP
OF LED DISPLAY WHILE PUMP IS RUNNING.
Rotor is turning, but is not
detected by rotor sensors. Do not try to repair pump. Return pump for
technical service.
NOTE: FOR ALL ALARM CONDITIONS,
PRESSING THE START/HOLD BUTTON WILL SILENCE THE AUDIBLE ALARM. THE VISIBLE
DISPLAY WILL CONTINUE TO FLASH UNTIL THE START/HOLD BUTTON IS PRESSED
AGAIN.
|
|
| Back to Table of Contents |
|
Cleaning
|
Important
User Information
CAUTION:
DO NOT IMMERSE PUMP OR POWER CORD IN WATER OR OTHER CLEANING SOLUTION;
CLEAN USING A DAMP (NOT WET) CLOTH OR SPONGE. FAILURE TO FOLLOW THE
CLEANING PROCEDURES DESCRIBED HEREIN COULD RESULT IN HAZARDS TO USERS. As
with any electrical device, care must be taken to prevent
liquid from entering the pump or the power cord receptacle to avoid
electrical shock hazard, fire hazard, or damage to electrical
components.
Should any of the following
events occur, DO NOT USE the pump until it has been properly cleaned and
serviced by personnel trained in servicing KangarooŽ pumps:
- wetting of the pump's power
cord or leakage into the pump interior during cleaning.
- spillage of large amounts of
formula onto the pump exterior or any spillage on the power cord.
General
Cleaning Directions
Cleaning of KangarooŽ 224 pumps
must be performed as follows:
CAUTION:
Disconnect pump from AC power source before cleaning. Upon completion of
cleaning, do not connect to AC power source until pump and power cord
are thoroughly dry.
- A mild detergent should be
used for general cleaning. If necessary, the pump may be cleaned with a
10:1 water and hypochlorite mixture, however, repeated cleaning with
this solution can damage the plastic pump case. Isopropyl alcohol
applied with a damp (not wet) cotton swab may be used for cleaning
difficult-to-reach areas,such as the drop sensor and inside the AC
receptacle, however, it should be used sparingly because repeat cleaning
may damage the case.
- DO NOT USE strong cleaners
such as Spray-Nine, Phisohex, Hibiclens, or Vesta-Syde because damage to
the pump case housing can result.
Directions For Cleaning Pump Housing
- Refer to General Cleaning
Directions before starting.
- Clean outside surface with a
damp (not wet) cloth or sponge and keep pump in upright position as much
as possible.
- Avoid excess moisture near
handle hinges.
- Avoid excess moisture near
pole clamp area.
- When cleaning around buzzer
opening on back of pump housing, do not let cleaning solution get inside
pump.
|
|
Directions For Cleaning Pump Power Cord
- Refer to General Cleaning
Directions before starting.
- Unless soiling is observed,
the power cord should not be cleaned.
- If cleaning of the power
cord is necessary, wipe the exterior surfaces of the wall plug of the power
cord with a cloth dampened with isopropyl alcohol.
CAUTION:
Avoid exposing power cord to excess moisture, as this can lead to an
electrical shock or fire hazard.
Directions For Cleaning Drop Detector
- Refer to General Cleaning
Directions before starting.
- Clean the drop sensors in the
upper drip chamber guide with a cotton swab dampened with isopropyl
alcohol.
- Be sure sensor areas are
clear.
Directions For Cleaning Rotor Assembly
- Refer to General Cleaning
Directions before starting.
- Loosen rotor set screw with a
0/64" allen wrench and gently pull rotor forward off shaft. After
removing rotor, avoid getting any moisture in the rotor shaft opening.
- Clean rollers thoroughly with
warm soapy water, or isopropyl alcohol if necessary.
- Be sure all parts of rotor are
completely dry before putting it back onto shaft.
- To replace rotor, align set
screw on hub of rotor with the flattened section of the output shaft.
Push rotor into place and
tighten set screw (Do not over tighten).
Preventive Maintenance
This pump must be periodically
serviced to assure proper functioning and safety. The recommended service
interval is 6 months, but in no case less than once each twelve (12)
months per JCAHO. Servicing must be performed by personnel trained in
servicing KangarooŽ pumps and may be done at the user's SPARROW REGIONAL
MEDICAL SUPPLY biomedical engineering department, an outside service, or
by Sherwood Factory Service. To arrange for Sherwood Factory Service, call
1-800-448-0190.
|
Back to Table of Contents |
| Troubleshooting Guide |
| Listed below are some
of the probable causes of alarm conditions and their
corrections.
|
| Symptom |
Probable Cause |
Corrections |
Flow Error Alarm (FLO Err) |
Feeding Container is
empty.
Feeding tube or feeding set
tubing occluded.
Drip chamber is
improperly placed in pump.
Drip chamber walls are coated
with feeding formula.
Sensors in upper drip
chamber guide are blocked.
|
Refill feeding container. If bag
has been in use for 24 hours, replace
Locate point of occlusion and
correct. Note: occlusion may occur upstream from pump.
- Feeding Container
- Feeding Pump Set (control
clamp closed)
- Feeding Tube-patency of
feeding tube should be checked
Check to make sure drip chamber
is properly "locked" in position is drip chamber.
Check to make sure formula is not
preventing detectors from proper operation. If formula cannot be removed
from inside walls of chamber via chamber manipulation, replace
set.
Check to make sure formula is not
preventing detectors from proper operation. If formula cannot be removed
from inside walls of chamber via chamber manipulation, replace
set.
|
| Low Battery Alarm (Lo bAt) |
Battery has been run
down below point of maintaining accuracy of pump.
|
Plug pump
power cord into outlet. |
| No Set (no SEt) |
Pump set
has been improperly placed into pump. |
Check to
make sure only KangarooŽ Pump Pet has been placed onto the pump. Check to
make sure that retainer clip on pump set has been properly positioned into
retainer lock on pump.
|
| System Error (SYs Err) |
Rotor is not turning while pump
is running.
Rotor is turning but is not
detected by rotor sensors.
|
Return pump
for technical service. |
| Back to Table of Contents |
Factory Service |
If the pump
malfunctions, please contact Sparrow Regional Medical Supply at 517.364.2115. The pump
can be repaired on site or sent back to Sherwood Medical.
|
Back to Table of Contents |
Specifications |
|
Type Infusion Device Volumetric
Pumping Mechanism Rotary Peristalic
Pump Set Required Sherwood Medical KangarooŽ Feeding Pump Administration
Sets
Operating Range
5-300 ml/hr increments
Memory of Settings When Pump is Off 24 hours for all
settings
Accuracy ą10% of selected flow rate
Occlusion Pressure 12 psi (82.7 kPa) Nominal
Alarms
- Flow Error
- Low Battery
- Hold Error
- No Set
- System Error
Battery
- Rechargeable
- Automatically charges when
pump is plugged into wall outlet. Recharge time for fully discharged
battery is approximately 15 hours.
- A new battery, when fully
charged, will operate for approximately 24 hours at 125 milliliters per
hour.
Dimensions Approximately 7.25" High x 5.5" Wide x
4.25" Deep
Weight Approximately 3.625 Ibs.
Case Material ABS Plastic -FIRE RESISTANT
Power Requirements 120V, 60 Hz, 1 Amp
|
Back to Table of Contents |
|
Warranty |
|
Sherwood Medical Company
(hereinafter referred to as Sherwood) warrants that each KangarooŽ Enteral
Feeding Pump will be free from defects in materials and workmanship under
normal use and service for a period of sixty (60) days from the date of
delivery by Sherwood to the first purchaser. If any such defect occurs
during the warranty period, the aforesaid purchaser should communicate
directly with Sherwood. If returned to Sherwood, Sherwood will arrange for
repairs or replacement within the terms of the warranty. The defective
pump should be promptly returned, properly packaged, postage prepaid. Loss
or damage in return shipment to Sherwood shall be at Purchaser's
risk.
In no event shall Sherwood be
liable for any incidental, indirect or consequential damages in connection
with the purchase or use of the pump. This warranty shall not apply to,
and Sherwood shall not be responsible for, any loss arising in connection
with the purchase or use of such pump which has been repaired by anyone
other than an authorized Sherwood representative or altered in any way as,
in Sherwood's judgement, to affect its stability or reliability or which
has been subject to misuse, negligence or accident, or which has or had
been used otherwise than in accordance with the instructions furnished by
Sherwood. Sherwood neither assumes nor authorizes any representative or
other person to assume for it any other liability in connection with the
sale of such pumps.
SHERWOOD DISCLAIMS ALL OTHER
WARRANTIES, EXPRESS OR IMPLIED, INCLUDING ANY IMPLIED WARRANTY OF
MERCHANTABILITY OR OF FITNESS FOR A PARTICULAR PURPOSE OF APPLICATION
OTHER THAN THOSE EXPRESSLY SET FORTH IN THE APPROPRIATE PRODUCT LABELING
OR USER INFORMATION MANUAL. IN NO EVENT WILL SHERWOOD BE LIABLE FOR ANY
INCIDENTAL, INDIRECT OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH THE
PURCHASE OR USE OF ITS PRODUCTS.
|
Kangaroo is a registered trademark of Sherwood Medical and is used with permission.
|
|